219
Changing the Filter Cube
23
Changing the Filter Cube
SMD filter cubes for the following dye combinations are available for the Leica SMD system:
Table 22: List of available SMD filter cubes
WARNING
Risk of permanent eye damage from laser radiation
Never change filter cubes during the scanning operation because laser
radiation can escape uncontrolled from the specimen area.
Dye
combination
Maximum
number of
external
detection
channels
Possible
excitation lines
Beam
splitter
Bandpass
I
Bandpass II
neutral
2
50/50 %
-
-
Polarization
2
Polarization
-
-
GFP wide / RFP
2
458, 470, 488, 561,
594
LP 560
500 – 550
607 – 683
Alexa 488 /
Alexa 633
2
458, 470, 488, 561,
594, 633
LP 620
500 – 550
647 – 703
CFP / YFP
2
405, 458, 514
LP 505
467 – 499
535 – 585
CFP/RFP
2
405, 458, 561, 594
LP 560
470 – 550
607 – 683
GFP narrow /
RFP
2
458, 470, 488, 543,
561, 594
LP 560
500 – 530
607 – 683
GFP / RFP
2
458, 470, 488, 561,
594
LP 560
500 – 550
607 – 683
FITC/TRITC
2
458, 470, 488, 543
LP 560
500 – 550
565 – 605
GFP/HcRed
2
458, 470, 488, 561
LP 560
500 – 550
581 – 654
Alexa 405/ cy2
2
405, 470, 488
LP 470
418 – 458
500-550
DAPI, Atto 647
2
405, 514, 543, 561,
594, 633, 640
LP 620
420 – 500
650-710
Cy2/Cy5
2
470, 561, 633, 640
LP 620
485 – 550
650-710
CFP 440/YFP
2
405, 440, 514
LP 505
465-500
535-585
CFP 440/HcRed
2
405, 440, 514, 543,
561
LP 560
465-500
581-654
empty
1
-
-
-
VIS
1
405
-
-
LP 430
DAPI
1
405
-
-
420 - 500
FITC
1
405, 470, 488
-
-
500 - 550
TRITC
1
514, 543
-
-
565 - 605
Cy5
1
562, 594, 633
-
-
650 - 710
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......