Available actions when Approvals=OIML
Table 10-6:
Function
Transmitter status
Unsecured
Secured
Connecting from ProLink II, ProLink III, the
Field Communicator, or a host
All supported connec-
tion types are available.
Physical security prevents access to the
RS-485 terminals, so neither RS-485 con-
nections nor service port connections
are available.HART/Bell 202 connections
can be used to read data from the trans-
mitter, but software security prevents
writing to the transmitter.
Configuration
Reading
Allowed
Allowed
Changing
Allowed
Not allowed
Calibration
Zero
Allowed
Not allowed
Density
Allowed
Not allowed
Temperature
Allowed
Not allowed
Smart Meter Verifica-
tion
Outputs set to Continue
Measuring
Allowed
Allowed
Outputs set to Fault
Allowed
Allowed
Outputs set to Last
Measured Value
Allowed
Not allowed
External pressure and
temperature data
Retrieved by polling
Allowed
Allowed
Written by Modbus or
HART host
Allowed
Not allowed
Outputs
mA output loop test
Allowed
Not allowed
mA output trim
Allowed
Not allowed
Frequency output loop
test
Allowed
Not allowed
Discrete output loop
test
Allowed
Not allowed
Process variables
All flow rates
Reported normally
Reported normally
Density
Reported normally
Reported normally
Temperature
Reported normally
Reported normally
Totalizers
Resetting
Allowed. If any single
totalizer is reset, all
other totalizers are re-
set automatically.
May or may not be allowed, depending
on configuration. Can be performed only
if flow rate is zero. If any single totalizer
is reset, all other totalizers are reset au-
tomatically.
Starting
Not allowed
N/A
Stopping
N/A
Not allowed
Inventories
Resetting
Allowed
Not allowed
Sensor simulation
Allowed
Not allowed
Operate the transmitter with the Weights & Measures application
Configuration and Use Manual
137
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...