6.4.3
Configure Discrete Output Fault Action
ProLink III
Device Tools > Configuration > Fault Processing
Field Communicator
•
Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO 1 Fault Action
•
Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO 2 Fault Action
Overview
Discrete Output Fault Action controls the behavior of the Discrete Output if the
transmitter encounters an internal fault condition.
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement
the fault action until the timeout has elapsed.
CAUTION!
Do not use Discrete Output Fault Action as a fault indicator. If you do, you may not be able to
distinguish a fault condition from a normal operating condition. If you want to use the Discrete
Output as a fault indicator, set Discrete Output Source to
Fault
and set Discrete Output Fault
Action to
None
.
Procedure
Set Discrete Output Fault Action as desired.
The default setting is
None
.
Related information
Fault indication with the Discrete Output
Options for Discrete Output Fault Action
Label
Discrete Output behavior
Upscale
•
Fault: Discrete Output is ON (site-specific voltage)
•
No fault: Discrete Output is controlled by its assign-
ment
Downscale
•
Fault: Discrete Output is OFF (0 V)
•
No fault: Discrete Output is controlled by its assign-
ment
None
(default)
Discrete Output is controlled by its assignment
Integrate the meter with the control system
96
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...