D3 or D3 and D4 density calibration using the Field Communicator
Figure 11-2:
D3 Calibration
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid
On-Line Menu >
Service Tools >
Maintenance >
Density Calibration
Dens Pt 3 T-Series
Calibration Method
executes
Enter density of D3 fluid
OK
Calibration in Progress
message
Density Calibration Complete
message
OK
Home
D4 Calibration
Fill sensor with D4 fluid
Dens Pt 4 T-Series
Calibration Method
executes
Enter density of D4 fluid
OK
Calibration in Progress
message
Density Calibration Complete
message
OK
Home
Done
Service Tools >
Maintenance >
Density Calibration
11.10
Perform temperature calibration
Temperature calibration establishes the relationship between the temperature of the
calibration fluids and the signal produced by the sensor.