Run an SMV test using the Field Communicator
1.
Navigate to the Smart Meter Verification menu:
•
Overview > Shortcuts > Meter Verification
•
Service Tools > Maintenance > Routine Maintenance > Meter Verification
2.
Choose Manual Verification.
3.
Choose Start.
4.
Set output behavior as desired, and press OK if prompted.
Option
Description
Continue Measuring
During the test, all outputs will continue to report their assigned proc-
ess variable. The test will run for approximately 90 seconds.
Outputs Held at Last
Value
During the test, all outputs will report the last measured value of their
assigned process variable. The test will run for approximately
140 seconds.
Outputs Held at Fault
During the test, all outputs will go to their configured fault action. The
test will run for approximately 140 seconds.
Test progress is displayed on the screen.
Postrequisites
View the test results and take any appropriate actions.
11.2.4
View test data
You can view the results of the current test. You can also view results from previous tests.
Important
You can view previous test results and see detailed test reports only if SMV is licensed.
The transmitter stores the following information about the previous twenty SMV tests:
•
Powered-on hours at the time of the test.
•
Test result (Pass, Fail, Abort).
•
Abort code, if applicable.
In addition, ProLink III provides a detailed test reporting and analysis framework. This
information is stored on the PC where ProLink III is installed for tests that were run only on
that PC. It includes:
•
Timestamp from the PC clock
•
Current flowmeter identification data
•
Current flow and density configuration parameters
•
Current zero values
Measurement support
144
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...