Understanding SMV results
When the SMV test is completed, the result is reported as Pass, Fail, or Abort. (Some tools
report the Fail result as
Advisory
instead.)
Pass
Smart Meter Verification does a statistical check between the factory baseline
value and the current Smart Meter Verification result. Pass indicates that the two
values are statistically the same.
Fail
The current SMV value is statistically different than the factory baseline value.
•
If the meter passes the second test, the first result can be ignored.
•
If the meter fails the second test, the flow tubes may be damaged. Call
customer support.
Abort
A problem occurred with the meter verification test (such as process instability)
or you stopped the test manually. See
description of each code, and possible actions you can take in response.
SMV abort codes
Table 11-3:
Code
Description
Recommended actions
1
User-initiated abort
None required. Wait 15 seconds before starting an-
other test.
3
Frequency drift
Ensure that temperature, flow, and density are sta-
ble, and rerun the test.
5
High drive gain
Ensure that flow is stable, minimize entrained gas,
and rerun the test.
8
Unstable flow
Check factors that could cause process instability,
then rerun the test. To maximize process stability:
•
Maintain a constant fluid pressure and tempera-
ture.
•
Avoid changes to fluid composition, such as
two-phase flow or settling.
•
Maintain a constant flow rate.
12
There is a fault present and
SMV cannot run
View the Alerts present on the device and take any
necessary actions to clear them.
13
No factory reference data for
meter verification test per-
formed on air
Contact customer service.
14
No factory reference data for
meter verification test per-
formed on water
Contact customer service.
15
No configuration data for
meter verification
Contact customer service.
Other
General abort
Repeat the test. If the test aborts again, contact
customer service.
Measurement support
146
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...