Problem
Possible causes
Recommended actions
Inaccurate flow rate
or batch total
•
Wiring problem
•
Inappropriate measurement unit
•
Incorrect flow calibration factor
•
Incorrect meter factor
•
Incorrect density calibration factors
•
Incorrect grounding
•
Two-phase flow
•
Problem with receiving device
•
Incorrect sensor zero
•
Check the wiring between the sensor and
the transmitter.
•
Verify that the measurement units are con-
figured correctly for your application.
•
Verify all of the characterization or calibra-
tion parameters. See the sensor tag or the
calibration sheet for your meter.
•
Perform a bucket test to verify batch to-
tals.
•
Zero the meter.
•
Check the grounding of all components.
•
Check for two-phase flow.
•
Verify the receiving device, and the wiring
between the transmitter and the receiving
device.
•
Check sensor coil resistance and for shorts
to case.
•
Replace the transmitter.
12.4
Density measurement problems
Problem
Possible causes
Recommended actions
Inaccurate density
reading
•
Problem with process fluid
•
Incorrect density calibration factors
•
Wiring problem
•
Incorrect grounding
•
Two-phase flow
•
Plugged or coated sensor tube
•
Incorrect sensor orientation
•
RTD failure
•
Physical characteristics of sensor have
changed
•
Check your process conditions against the
values reported by the device.
•
Ensure that all of the calibration parame-
ters have been entered correctly. See the
sensor tag or the calibration sheet for your
meter.
•
Check the wiring between the sensor and
the transmitter.
•
Check the grounding of all components.
•
Verify all of the characterization or calibra-
tion parameters. See the sensor tag or the
calibration sheet for your meter.
•
Check for two-phase flow.
•
If two sensors with similar frequency are
too near each other, separate them.
•
Purge the sensor tubes.
Troubleshooting
172
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...