Available actions when Approvals=NTEP
Table 10-5:
Function
Transmitter status
Unsecured
Secured
Connecting from ProLink II, ProLink III, the
Field Communicator, or a host
All supported connec-
tion types are availa-
ble.
Physical security prevents access to the
RS-485 terminals, so neither RS-485
connections nor service port connec-
tions are available.HART/Bell 202 con-
nections can be used to read data from
the transmitter, but software security
prevents writing to the transmitter.
Configuration
Reading
Allowed
Allowed
Changing
Allowed
Not allowed
Calibration
Zero
Allowed
Not allowed
Density
Allowed
Not allowed
Temperature
Allowed
Not allowed
Smart Meter Verification
Outputs set to Continue
Measuring
Allowed
Allowed
Outputs set to Fault
Allowed
Allowed
Outputs set to Last
Measured Value
Allowed
Not allowed
External pressure and
temperature data
Retrieved by polling
Allowed
Allowed
Written by Modbus or
HART host
Allowed
Not allowed
Testing and trimming out-
puts
mA output loop test
Allowed
Not allowed
mA output trim
Allowed
Not allowed
Frequency output loop
test
Not allowed
Not allowed
Discrete output loop
test
Allowed
Not allowed
Totalizers
Resetting
Allowed. If any single
totalizer is reset, all
other totalizers are re-
set automatically.
May or may not be allowed, depending
on configuration. Can be performed on-
ly if flow rate is zero. If any single totaliz-
er is reset, all other totalizers are reset
automatically.
Starting
Not allowed
N/A
Stopping
N/A
Not allowed
Inventories
Resetting
Not allowed
Not allowed
Sensor simulation
Allowed
Not allowed
Operate the transmitter with the Weights & Measures application
136
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...