Alert num-
ber
Alert title
Possible cause
Recommended actions
A104
Calibration in Pro-
gress
A calibration procedure is in proc-
ess.
•
Allow the procedure to complete.
•
For zero calibration, you may abort the
calibration, set Zero Time to a lower val-
ue, and restart the calibration.
A105
Two-Phase Flow
The line density is outside the user-
defined two-phase flow limits.
•
Check for two-phase flow.
•
Check the live density reading against
the upper and lower two-phase flow
limit settings.
A106
Burst Mode Enabled
HART burst mode is enabled.
•
No action required.
•
If desired, you can set Alert Severity
Level to
Ignore
.
A107
Power Reset Occur-
red
The transmitter has been restarted.
•
No action required.
•
If desired, you can set Alert Severity
Level to
Ignore
.
A108
Basic Event 1 On
The process has triggered Basic
Event 1.
•
No action required.
•
Review event configuration if you be-
lieve the event was triggered errone-
ously.
A109
Basic Event 2 On
The process has triggered Basic
Event 2.
•
No action required.
•
Review event configuration if you be-
lieve the event was triggered errone-
ously.
A110
Frequency Output
Saturated
The calculated Frequency Output is
outside the configured range.
•
Check the Frequency Output scaling.
•
Check process conditions. Actual condi-
tions may be outside the normal condi-
tions for which the output is config-
ured.
•
Check for air in the flow tubes, tubes
not filled, foreign material in the tubes,
coating in the tubes, or other process
problems.
•
Verify that the measurement units are
configured correctly for your applica-
tion.
•
Purge the sensor tubes.
A111
Frequency Output
Fixed
The Frequency Output has been
configured to send a constant val-
ue.
•
Cycling power to the meter or restart-
ing totalizers will restore the Frequency
Output to normal operation.
•
Check whether the output is in loop test
mode. If it is, unfix the output.
•
Check whether the output has been set
to a constant value via digital communi-
cation.
Troubleshooting
168
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...