6.3
Configure the Frequency Output
The Frequency Output is used to report a process variable. The Frequency Output
parameters control how the process variable is reported. Your transmitter may have zero,
one, or two Frequency Outputs, depending on the configuration of Channels B and C. If
both Channels B and C are configured as Frequency Outputs, they are electrically isolated
but not independent. You cannot configure them separately.
Important
Whenever you change a Frequency Output parameter, verify all other Frequency Output parameters
before returning the flowmeter to service. In some situations, the transmitter automatically loads a
set of stored values, and these values may not be appropriate for your application.
6.3.1
Configure Frequency Output Process Variable
ProLink III
Device Tools > Configuration > I/O > Outputs > Frequency Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Settings > Third
Variable
Overview
Frequency Output Process Variable controls the variable that is reported over the
frequency output.
Prerequisites
If you plan to configure the output to report volume flow, ensure that you have set Volume
Flow Type as desired:
Liquid
or
Gas Standard Volume
.
If you plan to configure an output to report a concentration measurement process
variable, ensure that the concentration measurement application is configured so that the
desired variable is available.
If you are using the HART variables, be aware that changing the configuration of Frequency
Output Process Variable will change the configuration of the HART Tertiary Variable (TV).
Procedure
Set Frequency Output Process Variable as desired.
The default setting is
Mass Flow Rate
.
Options for Frequency Output Process Variable
The transmitter provides a basic set of options for Frequency Output Process Variable, plus
several application-specific options. Different communications tools may use different
labels for the options.
Integrate the meter with the control system
86
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...