1.
Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions. During the power-up sequence, Alert A009 is active.
This alert should clear automatically when the power-up sequence is complete.
2.
Check the status LED on the transmitter.
Related information
View and acknowledge status alerts
2.2.1
Transmitter status reported by LED
Status LED states
Table 2-1:
LED state
Alarm condition
Description
Solid green
No alarm
Normal operation
Flashing yellow
No alarm
•
Zero calibration procedure is in progress
•
Loop test is in progress
Solid yellow
Low-severity alarm
Alarm condition that will not cause measure-
ment error (outputs continue to report proc-
ess data)
Solid red
High-severity alarm
Alarm condition that will cause measurement
error (outputs in fault)
2.3
Make a startup connection to the transmitter
To configure the transmitter, you must have an active connection from a communications
tool. Follow this procedure to make your first connection to the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in
the appropriate appendix. Use the default communications parameters shown in the
appendix.
Communications tool
Connection type to use
Instructions
ProLink III
Modbus/RS-485
Field Communicator
HART
Quick start
6
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...