Transmitter outputs and process data when Approvals=NTEP
(continued)
Table 10-3:
Function
Transmitter status
Unsecured
Secured
Frequency output behavior Inactive (produces no pulses), even under fault
conditions
Normal
Discrete output behavior
Normal
Normal
Process variables All flow rates
Reported as zero
Reported normally
Density
Reported normally
Reported normally
Temperature
Reported normally
Reported normally
Totalizers
Values
Do not increment or decrement
Increment normally
Inventories
Values
Do not increment or decrement
Increment normally
Transmitter outputs and process data when Approvals=OIML
Table 10-4:
Function
Transmitter status
Unsecured
Secured
Outputs
mA output behavior
Performs configured fault action
Normal
Frequency output behavior
Performs configured fault action
Normal
Discrete output behavior
Performs configured fault action
Normal
Process variables
All flow rates
Reported normally
Reported normally
Density
Reported normally
Reported normally
Temperature
Reported normally
Reported normally
Totalizers
Values
Do not increment or decrement
Increment normally
Inventories
Values
Do not increment or decrement
Increment normally
10.1.3
Effect of the Weights & Measures application on
operation and maintenance functions
When the Weights & Measures application is installed, specific operations and
maintenance activities may be disallowed. You can switch to unsecured mode, perform
the required actions, then switch back to secured mode.
Operate the transmitter with the Weights & Measures application
Configuration and Use Manual
135
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...