Procedure
Set Flow Damping to the value you want to use.
The default value is 0.8 seconds. The range depends on the core processor type and the
setting of Update Rate, as shown in the following table.
Update Rate setting
Damping range
Normal
0 to 51.2 seconds
Special
0 to 40.96 seconds
The value you enter is automatically rounded off to the nearest valid value. For example, if
the damping is currently set to 0.8 seconds, any value entered up to 1.2 seconds will be
rounded down to 0.8 seconds, and any value entered from 1.21 to 1.59 seconds will be
rounded up to 1.6 seconds.
Update Rate setting
Valid damping values
Normal
0.0, 0.2, 0.4, 0.8, 1.6, 3.2, 6.4, 12.8, 25.6, 51.2
Special
0.0, 0.04, 0.08, 0.16, 0.32, 0.64, 1.28, 2.56,
5.12, 10.24, 20.48, 40.96
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also
affects volume measurement for gas standard volume data. The transmitter calculates
volume data from the damped mass flow data.
Interaction between Flow Damping and mA Output
Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the
reported mass flow value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping
controls the rate of change reported via the mA Output. If mA Output Process Variable is
set to
Mass Flow Rate
, and both Flow Damping and mA Output Damping are set to non-zero
values, flow damping is applied first, and the added damping calculation is applied to the
result of the first calculation.
4.1.3
Configure Mass Flow Cutoff
ProLink III
Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Cutoff
Configure process measurement
20
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...