Valid values for Added Damping
(continued)
Table 6-7:
Setting of Update Rate
Process variable
Update rate
in effect
Valid values for Added Damping
Special
100 Hz variable (if assigned
to the mA Output)
100 Hz
0.0
,
0.04
,
0.12
,
0.30
,
0.64
,
1.32
,
2.6
,
5.4
,
11
,
22
,
44
,
88
,
176
,
350
100 Hz variable (if not as-
signed to the mA Output)
All other process variables
6.25 Hz
0.0
,
0.32
,
0.96
,
2.40
,
5.12
,
10.56
,
20.8
,
43.2
,
88
,
176
,
352
Interaction between mA Output Damping and process
variable damping
When mA Output Source is set to a flow rate variable, density, or temperature, mA Output
Damping interacts with Flow Damping, Density Damping, or Temperature Damping. If
multiple damping parameters are applicable, the effect of damping the process variable is
calculated first, and the mA Output damping calculation is applied to the result of that
calculation.
Example: Damping interaction
Configuration:
•
Flow Damping =
1
second
•
mA Output Source =
Mass Flow Rate
•
mA Output Damping =
2
seconds
Result: A change in the mass flow rate will be reflected in the mA Output over a time
period that is greater than 3 seconds. The exact time period is calculated by the
transmitter according to internal algorithms which are not configurable.
6.2.5
Configure mA Output Fault Action and mA Output Fault
Level
ProLink III
Device Tools > Configuration > Fault Processing
Field Communicator
•
Configure > Manual Setup > Inputs/Outputs > mA Output 1 > MA01 Fault Settings
•
Configure > Manual Setup > Inputs/Outputs > mA Output 2 > MA02 Fault Settings
Overview
mA Output Fault Action controls the behavior of the mA Output if the transmitter
encounters an internal fault condition.
Integrate the meter with the control system
84
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...