PVR-only HART process variables
(continued)
Table 6-25:
Process variable
Primary Varia-
ble (PV)
Secondary
Variable (SV)
Third Variable
(TV)
Fourth Varia-
ble (QV )
Volume Flow of Mix at Line
✓
✓
✓
Volume Total Of Mix @ Line
✓
Water Total @ Line
✓
TMR-only HART process variables
Table 6-26:
Process variable
Primary Varia-
ble (PV)
Secondary
Variable (SV)
Third Variable
(TV)
Fourth Varia-
ble (QV )
Remediated Mass Flow
✓
✓
✓
Remediated Mass Total
✓
Remediated Mass Inventory
✓
PVR- and TBR-only HART process variables
Table 6-27:
Process variable
Primary Varia-
ble (PV)
Secondary
Variable (SV)
Third Variable
(TV)
Fourth Varia-
ble (QV )
Unremediated Density
✓
✓
✓
PVR, TBR, and TMR HART process variables
Table 6-28:
Process variable
Primary Varia-
ble (PV)
Secondary
Variable (SV)
Third Variable
(TV)
Fourth Varia-
ble (QV )
Total Remediated Time
✓
Interaction of HART variables and transmitter outputs
The HART variables are automatically reported through specific transmitter outputs. They
may also be reported through HART burst mode, if enabled on your transmitter.
HART variables and transmitter outputs
Table 6-29:
HART variable
Reported via
Comments
Primary Variable (PV)
Primary mA output
If one assignment is changed, the other is changed auto-
matically, and vice versa.
Secondary Variable (SV)
Secondary mA Output (if
present on your transmitter)
If one assignment is changed, the other is changed auto-
matically, and vice versa. If your transmitter is not config-
ured for a secondary mA Output, the SV must be config-
ured directly, and the value of the SV is available only via
digital communications.
Integrate the meter with the control system
108
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...