Option
Description
Informa-
tional
Actions when fault is detected:
•
The alert is posted to the Alert List.
•
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
•
The status LED (if available) returns to green and may or may not flash.
Ignore
No action
Status alerts and options for Status Alert Severity
Status alerts and Status Alert Severity
Table 5-2:
Alert code
Status message
Default severity Notes
Configurable?
A001
EEPROM Error
Fault
No
A002
RAM Error
Fault
No
A003
No Sensor Response
Fault
Yes
A004
Temperature Overrange
Fault
No
A005
Mass Flow Rate Overrange
Fault
Yes
A006
Characterization Required
Fault
Yes
A008
Density Overrange
Fault
Yes
A009
Transmitter Initializing/
Warming Up
Fault
Yes
A010
Calibration Failure
Fault
No
A011
Zero Calibration Failed:
Low
Fault
Yes
A012
Zero Calibration Failed:
High
Fault
Yes
A013
Zero Calibration Failed:
Unstable
Fault
Yes
A014
Transmitter Failure
Fault
No
A016
Sensor RTD Failure
Fault
Yes
A017
T-Series RTD Failure
Fault
Yes
A018
EEPROM Error (Transmit-
ter)
Fault
No
A019
RAM Error (Transmitter)
Fault
No
A020
No Flow Cal Value
Fault
Yes
A021
Incorrect Sensor Type (K1)
Fault
No
A022
Configuration Database
Corrupt (Core Processor)
Fault
Applies only to flowmeters with the
standard core processor.
No
Configure device options and preferences
68
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...