Mass Flow Cutoff:
10
g/sec
Result:
•
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
-
The mA Output will report zero flow.
-
The Frequency Output will report the actual flow rate, and the actual flow rate
will be used in all internal processing.
•
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0
will be used in all internal processing.
4.2
Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured
and reported.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time.
Choose one or the other.
Note
If you need to switch from gas standard volume to liquid volume, polling for base density will
automatically be disabled.
4.2.1
Configure Volume Flow Type for liquid applications
ProLink III
Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Gas Standard Volume > Volume Flow Type
> Liquid
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement will
be used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type
to
Liquid
if you are using any of the following applications:
•
Petroleum measurement
•
Concentration measurement
•
Fuel consumption
Configure process measurement
22
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...