Overview
The Standard Density of Gas value is the gas density at standard reference conditions. Use
it to convert the measured mass flow data to volume flow at reference conditions.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for
Standard Density of Gas.
Procedure
From the Source field, choose the method to supply gas base density data and perform the
required setup.
Option
Description
Fixed Value or Digital
Communications
A host writes gas base density data to the meter at appropriate intervals.
Continue to
Configure fixed value or digital communications
.
Poll for external value The meter polls an external HART device for gas base density data in order
to then compute gas standard volume from the mass flow and gas base
density.
Continue to
.
Configure fixed value or digital communications
Prerequisites
Procedure
1.
Set Standard Density of Gas to the standard reference density of the gas you are
measuring.
Note
ProLink III provides a guided method that you can use to calculate your gas base density, if
you do not know it.
2.
Poll for external value
Prerequisites
Procedure
1.
Set Polling Slot to an available slot.
Configure process measurement
28
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...