Current process values for mass flow rate, volume flow rate, density, temperature,
and external pressure
•
Customer and test descriptions (if entered by the user)
You can use ProLink III to run a test that displays a test result chart and a test report at the
completion of the test. On-screen directions are provided to manipulate the test data or
export the data to a CSV file for offline analysis.
View test result data using ProLink III
Prerequisites
You can view test result data only if your SMV is licensed and only for tests that were run on
the PC you are currently using.
Procedure
1.
Choose Device Tools > Diagnostics > Meter Verification and click Previous Test
Results.
The chart shows test results for all tests stored in the ProLink III database.
2.
(Optional) Click Next to view and print a test report.
3.
(Optional) Click Export Data to CSV File to save the data to a file on your PC.
View test result data using the Field Communicator
Prerequisites
You can view test result data only if your SMV is licensed.
Procedure
1.
Navigate to the Smart Meter Verification menu:
•
Overview > Shortcuts > Meter Verification
•
Service Tools > Maintenance > Routine Maintenance > Meter Verification
2.
(Optional) If the Field Communicator database is out of date, choose Upload Results
Data from Device.
3.
To view data from the most recent test, choose Most Recent Test Results.
4.
To view data for all tests in the Field Communicator database:
a. Press Show Results Table.
Data from the most recent test is displayed.
b. Press OK to scroll through data from previous tests.
c. To exit the results table, press Abort.
Measurement support
Configuration and Use Manual
145
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...