Alert num-
ber
Alert title
Possible cause
Recommended actions
A021
Transmitter/Sensor/
Software Mismatch
The configured board type does not
match the physical board, or the
configured sensor type does not
match the physical sensor.
•
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
•
Ensure that the correct board is instal-
led.
•
Verify the setting of the Sensor Type pa-
rameter.
•
If Sensor Type=
Curved Tube
, ensure that
no parameters specific to
Straight Tube
have been set.
A027
Security Breach
The transmitter has been switched
from secure mode to unsecure
mode. In unsecured mode, the
transmitter configuration can be
changed.
The HART device ID is set to zero.
•
Check the HART device ID. For custody
transfer or MID applications, it must be
non-zero. If it is set to 0, contact the
factory.
•
Set the transmitter to
secure
to clear the
alert.
Note that additional procedures may be
required to return the device to secured
mode.
•
If the HART device ID is set to zero, en-
ter any nonzero value using ProLink III.
For example, enter the transmitter seri-
al number.
A030
Incorrect Board Type The loaded software is not compati-
ble with the programmed board
type.
•
Contact customer support.
A032
Meter Verification in
Progress: Outputs to
Fault
A meter verification test is in pro-
gress, with outputs set to
Fault
.
•
Allow the procedure to complete.
A033
Insufficient Pickoff
Signal
The signal from the sensor pick-
off(s) is insufficient. This suggests
that the sensor tubes or vibrating
elements are not vibrating. This
alert often occurs in conjunction
with Alert 102.
•
Check for air in the flow tubes, tubes
not filled, foreign material in the tubes,
coating in the tubes, or other process
problems.
•
Check for foreign material in the proc-
ess gas or fluid, coating, or other proc-
ess problems.
•
Check for fluid separation by monitor-
ing the density value and comparing
the results against expected density val-
ues.
•
Ensure that the sensor orientation is ap-
propriate for your application. Settling
from a two-phase or three-phase fluid
can cause this alert even if the flow
tubes are full.
Troubleshooting
166
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...