If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and
0 will be used in all internal processing.
4.3
Configure GSV flow measurement
The gas standard volume (GSV) flow measurement parameters control how volume flow is
measured and reported in a gas application.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time.
Choose one or the other.
4.3.1
Configure Volume Flow Type for gas applications
ProLink III
Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Standard Gas
Volume
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement is
used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type
to
Liquid
if you are using any of the following applications:
•
Petroleum measurement
•
Concentration measurement
•
Fuel consumption
•
Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to
Gas Standard Volume
.
4.3.2
Configure Standard Density of Gas
ProLink III
Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Gas Ref Density
Configure process measurement
Configuration and Use Manual
27
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...