a. Open the Petroleum Measurement Source menu and select the API table
number.
Depending on your choice, you may be prompted to enter a reference
temperature or a thermal expansion coefficient.
b. Enter the API table letter.
These two parameters uniquely specify the API table.
3.
Determine how the transmitter will obtain temperature data for the petroleum
measurement calculations, and perform the required setup.
Option
Setup
Temperature data
from the sensor
a. Choose Online > Configure > Manual Setup > Measurements >
External Pressure/Temperature > Temperature .
b. Set External Temperature to
Disabled
.
A user-configured
static temperature
value
a. Choose Online > Configure > Manual Setup > Measurements >
External Pressure/Temperature > Temperature .
b. Set External Temperature to
Enabled
.
c. Set Correction Temperature to the value to be used.
Polling for tempera-
ture
a. Ensure that the primary mA output has been wired to support
HART polling.
b. Choose Online > Configure > Manual Setup > Measurements >
External Pressure/Temperature > Temperature .
c. Set External Temperature to
Enabled
.
d. Choose External Polling.
e. Set Poll Control to
Poll As Primary
or
Poll as Secondary
.
f. Determine whether you will use Polling Slot 1 or Polling Slot 2.
g. For the chosen slot, set Ext Dev Tag to the HART tag of the exter-
nal temperature device.
h. For the chosen slot, set Polled Variable to
Temperature
.
Tip
•
Poll as Primary
: No other HART masters will be on the network.
•
Poll as Secondary
: Other HART masters will be on the network.
The Field Communicator is not a HART master.
Configure process measurement
48
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...