Transmitter default values and ranges
(continued)
Table C-1:
Type
Parameter
Default
Range
Comments
Special units
Base mass unit
g
Base mass time
sec
Mass flow conversion factor
1
Base volume unit
L
Base volume time
sec
Volume flow conversion factor
1
Variable map-
ping
Primary variable
Mass flow
Secondary variable
Density
Tertiary variable
Mass flow
Quaternary variable
Volume flow
mA Output 1
Primary variable
Mass flow
LRV
–200.00000
g/s
URV
200.00000 g/s
AO cutoff
0.00000 g/s
AO added damping
0.00000 sec
The user-entered value is correc-
ted down to the nearest lower
value in a list of preset values.
LSL
–200 g/s
Read-only.
LSL is calculated based on the
sensor size and characterization
parameters.
USL
200 g/s
Read only.
USL is calculated based on the
sensor size and characterization
parameters.
MinSpan
0.3 g/s
Read-only.
Fault action
Downscale
AO fault level – downscale
2.0 mA
1.0 – 3.6 mA
AO fault level – upscale
22 mA
21.0 – 24.0 mA
Last measured value timeout
0.00 sec
mA Output 2
Secondary variable
Density
LRV
0.00 g/cm3
URV
10.00 g/cm3
AO cutoff
Not-A-Number
AO added damping
0.00000 sec
Default values and ranges
Configuration and Use Manual
215
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...