Option
Description
Normal (Legacy)
Transmitter calculates process variables at the standard speed. This op-
tion is selected if this parameter was configured on an earlier version of
ProLink III software.
Special (Legacy)
Transmitter calculates process variables at a faster speed. This option is
selected if this parameter was configured on an earlier version of ProLink
III software.
Normal - Optimal Filtering
Transmitter calculates process variables at standard filtering and speed.
Low Filtering - Fastest Re-
sponse
Transmitter calculates process variables at the fastest speed.
High Filtering - Smoothest
Output
Transmitter calculates process variables at the smoothest (least noisy)
response to changes in the process.
Service
For factory use only.
5.2
Configure alert handling
The alert handling parameters control the transmitter’s response to process and device
conditions.
5.2.1
Configure Fault Timeout
ProLink III
Device Tools > Configuration > Fault Processing
Field Communicator Configure > Alert Setup > Alert Severity > Fault Timeout
Overview
Fault Timeout controls the delay before fault actions are performed.
Restriction
Fault Timeout is applied only to the following alerts (listed by Status Alert Code): A003, A004, A005,
A008, A016, A017, A033. For all other alerts, fault actions are performed as soon as the alert is
detected.
Procedure
Set Fault Timeout as desired.
The default value is 0 seconds. The range is 0 to 60 seconds.
If you set Fault Timeout to 0, fault actions are performed as soon as the alert condition is
detected.
The fault timeout period begins when the transmitter detects an alert condition. During
the fault timeout period, the transmitter continues to report its last valid measurements.
Configure device options and preferences
66
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...