f. Enable or disable the high and low limit alarms for temperature and density, as
desired.
Restriction
The high and low limit alarms require the enhanced core processor.
Example: If Alarm Limit is set to 5%, the high-temperature extrapolation alert is
enabled, and the matrix is built for a temperature range of 40 °F to 80 °F, an
extrapolation alarm will be posted if process temperature goes above 82 °F
6.
Select the label that will be used for the concentration unit.
a. Click Next.
b. On the Concentration Measurement page, set Concentration Units to the
desired label.
c. Set Concentration Units to the desired label.
d. If you set Units to
Special
, enter the custom label.
e. Click Finish.
7.
Determine how the transmitter will obtain temperature data for the concentration
measurement calculations, and perform the required setup.
Option
Setup
Temperature data
from the sensor
a. Choose Online > Configure > Manual Setup > Measurements .
b. Click External Inputs.
c. Click Next.
d. Disable External Temperature.
A user-configured
static temperature
value
a. Choose Online > Configure > Manual Setup > Measurements .
b. Click External Inputs.
c. Click Next.
d. Enable External Temperature.
e. Set Correction Temperature to the value to be used.
Configure process measurement
54
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...