11.10.2
Perform temperature calibration using the
Field Communicator
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Next
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration Complete
Temperature Slope calibration
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Done
Service Tools >
Maintenance >
Calibration >
Temperature Calibration
Offset
Calibration in Progress
Next
Enter temperature of high-
temperature fluid
Next
Calibration Complete
Service Tools >
Maintenance >
Calibration >
Temperature Calibration
Slope
Calibration in Progress
Finish
Measurement support
160
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs