12.1
Status LED states
The status LED on the transmitter indicates whether or not alerts are active. If alerts are
active, view the alert list to identify the alerts, then take appropriate action to correct the
alert condition.
Status LED states
Table 12-1:
LED state
Alarm condition
Description
Solid green
No alarm
Normal operation
Flashing yellow
No alarm
•
Zero calibration procedure is in progress
•
Loop test is in progress
Solid yellow
Low-severity alarm
Alarm condition that will not cause measure-
ment error (outputs continue to report proc-
ess data)
Solid red
High-severity alarm
Alarm condition that will cause measurement
error (outputs in fault)
12.2
Status alerts, causes, and recommendations
Alert num-
ber
Alert title
Possible cause
Recommended actions
A001
EEPROM Error
The core processor has experienced
a memory error.
•
Cycle power to the meter.
•
Replace the core processor.
•
Contact customer support.
A002
RAM Error
The core processor has experienced
a memory error.
•
Cycle power to the meter.
•
Replace the core processor.
•
Contact customer support.
A003
No Sensor Response The transmitter is not receiving one
or more basic electrical signals from
the sensor.
This alert often occurs in conjunc-
tion with Alert 102.
•
Check the drive gain and the pickoff
voltage.
•
Check the wiring between the sensor
and the transmitter.
•
Verify that internal wiring is secure and
that there are no internal electrical
problems.
•
Check the integrity of the sensor tubes.
•
Perform sensor coil resistance checks.
Troubleshooting
162
Micro Motion Model 2500 Transmitters with Configurable Input/Outputs
Summary of Contents for Micro Motion 2500
Page 8: ...Contents vi Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 10: ...Getting started 2 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 18: ...Quick start 10 Micro Motion Model 2500 Transmitters with Configurable Input Outputs ...
Page 237: ...NE 53 history Configuration and Use Manual 229 ...