It is critical that each tubing set is correctly connected from the
syringe valve, through the light shield, to the injector tip
holder.
Otherwise, injected fluid may miss the intended
well.
8. Remove the tubing feed-through cover from the top of the reader (2 screws). Store
the cover and screws with the shipping hardware in case the reader needs to be
shipped again.
9. Thread the injector tip holder, with outlet tubing connected to both ports, through the
hole in the top of the reader.
10. Open the reader's top access door, and, holding the injector tip holder by the tab,
insert the injector tips into the appropriate holes inside the reader.
A magnet helps guide the tips into place and secures them in the reader.
Figure 2-3: Injector tip holder in its socket
11. Seat the light shield in the reader's cover and finger-tighten the thumbscrews to
secure it.
14 | Chapter 2: Installation
BioTek Instruments, Inc.