Handling Filters and Mirrors
Filters and mirrors are stored in a filter cube as described in the overview section starting
on page
. If you run different types of fluorescence and luminescence assays, you can
replace the entire filter cube with a different one; this is the BioTek-recommended option.
Alternatively, you can install different filters or mirrors in the cube; this section describes
how to do this. Contact BioTek for more information on purchasing additional cubes, filters,
and mirrors.
Change a Filter or Mirror
Important!
After changing the contents of a filter cube, be sure to
update Gen5 with the new filter and mirror configuration and then send
the information to the reader. It is critical that the Gen5 Filter Cube
table reflect the actual location and characteristics of the filters and
mirrors in the installed filter cube. See instructions on page
.
Gather the following tools:
l
7/64" hex key
l
Lens paper
l
Cotton swab
l
Linen or cloth gloves
To remove a filter, plug, or mirror:
i
Handle with care. The mirrors are seated on a shelf in the bottom of the cube and are
not secured in place.
1. Remove the filter cube as instructed on page
2. Set the cube on a flat work surface (do not hold it in your hand). When you remove
the filter cube's top in step 4, the mirrors will fall out if the cube is not on a stable,
flat surface.
3. Using a 7/64" hex key, remove the screw and washer located between the two
emission filter positions (shown in
).
Synergy H1 Operator's Manual
Handling Filters and Mirrors | 49
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...