Synergy H1 M_FI_T_SF.prt
and
Synergy H1 M_FI_B_SF.prt
Parameter
Setting
Plate Type
"Costar 96 black opaque" (Top)
"Greiner SensoPlate" (#655892) (Bottom)
Read Step 1
Kinetic loop
Run time 0:01:00, Interval 0:00:04 (16 reads)
Detection Method
Fluorescence
Read Type
Endpoint / Kinetic
Optics Type
Monochromators
Step Label
"Sensitivity Read"
Read Well
D7
Wavelengths
1
Excitation
485 nm
Emission
528 nm
Optics Position
Top (or Bottom)
Gain
Auto, Scale to High Wells, D7, 50000
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
100
Lamp Energy
Low (faster)
Read Height
7.00 mm (Top only)
Read Step 2
Kinetic loop
Run time 0:01:30, Interval 0:00:06 (16 reads)
Detection Method
Fluorescence
Read Type
Endpoint / Kinetic
Optics Type
Monochromators
Step Label
"Sensitivity Read Buffer"
Read Well
C9, D9, E9
Wavelengths
1
Excitation
485 nm
Emission
528 nm
Optics Position
Top (or Bottom)
Gain
Auto, Use first filter set gain from FIRST Read
Step
Read Speed
Normal
Delay after plate movement
350 msec