Parameter
Setting
Measurements per data point
100
Lamp Energy
Low (faster)
Read Height
7.00 mm (Top only)
Read Step 3
Detection Method
Fluorescence intensity
Read Type
Endpoint / Kinetic
Optics Type
Filters
Step Label
"Corners Read"
Read well
Full Plate (to support Gen5 v1.11)
Filter Set
1 ("Green")
Excitation
485 nm
Emission
528 nm
Optics Position
Top (or Bottom)
Gain
Auto, Scale to High Wells, A3, 50000
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
100
Lamp Energy
Low (faster)
Read Height
7.00 mm (Top only)
Read Step 4
Detection Method
Fluorescence intensity
Read Type
Endpoint / Kinetic
Optics Type
Filters
Step Label
"Linearity Read"
Read well
C1-F5
Filter Set
1 ("Green")
Excitation
485 nm
Emission
528 nm
Optics Position
Top (or Bottom)
Gain
Auto, Scale to High Wells, C1, 50000
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
100
Lamp Energy
Low (faster)