of the green test dye solution.
l
Carefully set the plate aside.
9. Close the experiment without saving it.
If you are not using a BioTek absorbance reader, read the plate using the
parameters described in the table above. Perform the calculations and
dettermine pass/fail according to the instructions on page
.
10. If you are using a BioTek absorbance reader, configure Gen5 to communicate with
the reader.
11. Create an experiment based on the
Other Reader
protocol and read the plate.
12. When the experiment is complete, save the file with an identifying name.
13. Remove the plate from the carrier and set it aside.
14. Repeat the procedure using the
Synergy H1 Disp 2 Test No Read
protocol and a
new microplate.
15. When the tests are complete:
l
Prime both dispensers with at least 5000 µL of deionized water to flush out the dye
solution.
l
Refer to the instructions on page
to perform calculations and determine
pass/fail.
l
Test descriptions are provided on page
.
138 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.