Empty/Clean the Tip Priming Trough
Applies only to models equipped with injectors
The tip priming trough is a removable cup located in the rear pocket of the microplate
carrier, used for performing the tip prime. The trough holds about 1.5 mL of liquid and
must be periodically emptied and cleaned.
1. Extend the microplate carrier and carefully remove the tip priming trough from its
pocket in the left rear of the carrier.
2. Wash the trough in hot, soapy water. Use a small brush to clean in the corners.
3. Rinse the trough thoroughly and allow it to dry completely.
4. Replace the trough in the microplate carrier.
At the start of an experiment that requires dispensing, Gen5 prompts the user to empty
the tip prime trough.
Clean the Priming Plate
Applies only to models equipped with injectors
Clean the priming plate regularly to prevent bacteria growth and residue buildup. Wash the
plate in hot, soapy water, using a small brush to clean in the corners. Rinse thoroughly and
allow it to dry completely.
Clean the Dispense Tubes and Injectors
Applies only to models equipped with injectors
The dispense tubes and injectors require routine cleaning, at least quarterly and possibly
more frequently depending on the type of fluids dispensed.
Required Materials
Protective gloves and safety glasses
Mild detergent
Clean, lint-free cotton cloths
Deionized or distilled water
Stylus (affixed to the rear of the dispense module or reader) (BTI #2872304)
Synergy H1 Operator's Manual
Empty/Clean the Tip Priming Trough | 63
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...