Parameter
Setting
Gain
Auto, Scale to High Wells, C1, 80000
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
100
Lamp Energy
Low (faster)
Read Height
7.00 mm
Results Analysis
The Fluorescence Liquid Test procedures begin on page
Corners Test
1. Calculate the Mean of the 12 "corner" wells (A1–A3, A10–A12, H1–H3, and H10–
H12).
2. Calculate the Standard Deviation of the same 12 wells.
3. Calculate the %CV: (Standard Deviation/Mean)*100
The %CV must be
<3.0
to pass.
Sensitivity Test
1. Calculate the Mean and Standard Deviation of the 16 reads for each of the buffer
wells (C9, D9, E9).
2. Among the three buffer wells, find the Median Standard Deviation and
corresponding Mean.
3. Calculate the Mean for the 16 reads of the SF (or MUB) Concentration well (D7).
4. Calculate the Signal-to-Noise Ratio (SNR) using the Mean SF (or MUB)
Concentration, Buffer Median STD with its corresponding Buffer Mean:
(<SF or MUB>Mean-BufferMean)/(3*BufferSTD)
5. Calculate the Detection Limit:
Sodium Fluorescein:
Using the known concentration value of SF and the
calculated SNR: 1000/SNR
o
Filter-based system, top optics:
The Detection Limit must be <=10 pM (2
pg/mL) to pass.
o
Monochromator-based system, top/bottom optics:
The Detection Limit must be
<=20 pM (3.76 pg/mL) to pass.
Methylumbelliferone:
Using the known concentration value of MUB and the
calculated SNR: 17.6/SNR
98 | Chapter 7: Instrument Qualification Process
BioTek Instruments, Inc.
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...