Upper Right x= 8912 y= 6612
Delta 1 -856 - -852= -4
Delta 2 8912 - 8916= -4
Delta 3 6612 - 6612= +0
Delta 4 404 - 400= +4
Carrier - Top Filter Fluorescence
Upper Left x=-3672 y= 6612
Lower Left x=-3668 y= 404
Lower Right x= 6096 y= 396
Upper Right x= 6096 y= 6612
Delta 1 -3672 --3668= -4
Delta 2 6096 - 6096= +0
Delta 3 6612 - 6612= +0
Delta 4 396 - 404= -8
Carrier - Injectors
Upper Left x= 2196 y= 6616
Lower Left x= 2196 y= 404
Lower Right x=11964 y= 396
Upper Right x=11964 y= 6612
Delta 1 2196 - 2196= +0
Delta 2 11964 -11964= +0
Delta 3 6612 - 6616= -4
Delta 4 396 - 404= -8
Carrier - Test Sensors
Middle Sensor x=20628
Tested 20636
Delta +8
Probe Height 25.99 mm
Filter/Mirror Slider 4424
Mono Probe Changer 3072
Backlash 44
Excitation Monochromator
Absorbance B=-0.00241697 C=+0.50917959
305LP Edge +779.17
Tested +780.18
Emission Monochromator
Top Fluorescence B=-0.00231930 C=+1.28592360
Bottom Fluorescence B=-0.00029174 C=-0.21822651
INCUBATION
Temperature Setpoint: 0.0 Current Average: 26.9 A/D Test: PASS
Zone 1: 27.1 Min: 27.1 Max: 27.2 Range: PASS Thermistor: PASS
Zone 2: 27.0 Min: 27.0 Max: 27.1 Range: PASS Thermistor: PASS
Zone 3: 26.8 Min: 26.8 Max: 26.8 Range: PASS Thermistor: PASS
Zone 4: 26.7 Min: 26.6 Max: 26.7 Range: PASS Thermistor: PASS
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...