Gen5 protocols listed below (as applicable for your reader model) and described in
detail under
starting on page
For the
filter-based
fluorescence system:
Synergy H1_FI_T_SF.prt
Corners, Sensitivity, Linearity tests, Top
optics, Sodium Fluorescein (SF)
Synergy H1_FI_T_MUB.prt
Sensitivity and Linearity tests, Top optics,
Methylumbelliferone (alternate/
supplemental test for Top optics)
Synergy H1_FP.prt
Fluorescence Polarization (FP) test
Synergy H1_TRF.prt
Time-Resolved Fluorescence (TRF) test
For the
monochromator-based
fluorescence system:
Synergy H1_M_FI_B_SF.prt
Corners, Sensitivity, Linearity tests, Bottom
optics, Sodium Fluorescein (SF)
Synergy H1_M_FI_T_SF.prt
Corners, Sensitivity, Linearity tests, Top
optics, Sodium Fluorescein (SF)
Synergy H1_M_FI_T_MUB.prt
Sensitivity and Linearity tests, Top optics,
Methylumbelliferone (alternate/
supplemental test for Top optics)
l
Filter set definitions, as applicable ("Green"=SF, "Blue"=MUB):
Filter Set Name Green
Excitation: Band Pass, 485/20
Mirror: Dichroic, Top 510 nm (440/505, 515/640)
Emission: Band Pass, 528/20
Filter Set Name Blue
Excitation: Band Pass, 360/40
Mirror: Dichroic, Top 400 nm (320/390, 410/800)
Emission: Band Pass, 460/40
Filter Set Name TRF
Excitation: Band Pass, 360/40
Mirror: Dichroic, Top 400 nm (320/390, 410/800)
Emission: Band Pass, 620/40
Filter Set Name FP
(enable 'Fluorescence Polarization Cube')
Excitation: Band Pass, 485/20
Mirror: Dichroic, Top 510 nm (440/505, 515/640)
Emission: Band Pass, 528/20
122 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...