Parameter
Setting
Excitation
360/40 nm
Emission
620/40 nm
Optics Position
Top 400 nm
Gain
Auto, Scale to High Wells, A8, 50000
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
20
Lamp Energy
Low (faster)
Delay before collecting data
300 µsec
Data collection time
1000 µsec
Read Height
7.00 mm
Read Step 2
Kinetic loop
Run time 0:00:45, Interval 0:00:03 (16 reads)
Detection Method
Time-resolved fluorescence
Read Type
Endpoint / Kinetic
Optics Type
Filters
Step Label
"Sensitivity Read Buffer"
Read well
A6, B6, C6
Filter Set
1 ("TRF")
Excitation
360/40 nm
Emission
620/40 nm
Optics Position
Top 400 nm
Gain
Auto, Use first filter set sensitivity from
FIRST Read Step
Read Speed
Normal
Delay after plate movement
350 msec
Measurements per data point
20
Lamp Energy
Low (faster)
Delay before collecting data
300 µsec
Data collection time
1000 µsec
Read Height
7.00 mm
94 | Chapter 7: Instrument Qualification Process