Pass/Fail Criteria
l
If the reader is equipped with the low-noise PMT, the detection limit must be <=75
amol to pass.
l
If the reader is equipped with the red-shifted PMT, the detection limit must be <=500
amol to pass.
If you do not know which PMT is installed (#49984=low-noise PMT;
#49721=red-shifted PMT), please contact BioTek TAC.
Troubleshooting
If a test fails, try the suggestions below. If a test continues to fail, contact BioTek's
Technical Assistance Center (TAC).
l
Ensure that the reading is performed through a hole in the filter cube, not through a
glass filter.
l
Verify that the filter cube definition in Gen5 matches the physical item.
l
The optical probe(s)
may
need to be cleaned; contact BioTek TAC for guidance.
86 | Chapter 7: Instrument Qualification Process
BioTek Instruments, Inc.