Time-Resolved Fluorescence (TRF) Test:
l
Pipette 200 μL of deionized water (DI) into wells A6, B6, C6.
l
If you have not already done so, shake the vial of 20 pM europium suspension
vigorously for 30 seconds prior to pipetting. Alternatively, sonicate or vortex the vial.
l
Pipette 200 μL of the 20 pM europium suspension (Eu) into well A8.
1
2
3
4
5
6
7
8
9
10
11
12
A
DI
Eu
B
DI
C
DI
D
E
F
G
H
Synergy H1 Operator's Manual
Fluorescence Liquid Tests | 129