Test Procedure
1. If you have not already done so, prepare the test solution; see instructions on page
.
Refer to the pipette map on the next page for the remaining steps.
2. Perform the Sensitivity/Linearity tests using the Top optics of the filter-based
fluorescence system:
l
Pipette the solutions into a new 96-well solid black or glass-bottom plate.
l
Create an experiment based on
Synergy H1_FI_T_MUB.prt
and read the plate.
3. If your reader is equipped with the monochromator-based fluorescence system,
perform the Sensitivity/Linearity tests for that system:
l
Create an experiment based on
Synergy H1_M_FI_T_MUB.prt
and read the
plate.
4. Save the experiments. Refer to the instructions starting on page
to perform
calculations and determine pass/fail.
l
Troubleshooting tips are provided on page
.
l
Test descriptions are provided on page
.
132 | Chapter 8: Instrument Qualification Procedures
BioTek Instruments, Inc.
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...