Synergy H1 Abs Test 2.prt
Parameter
Setting
Plate Type
96 WELL PLATE
Shake Step
Linear, 4 minutes, default frequency
Two Read Steps
Kinetic loop (one per Read step)
Set a Run Time/Interval combination to read the
plate five times with minimal delay
Detection Method
Absorbance
Read Type
Endpoint
Optics Type
Monochromators
Step labels
First Read step: "Normal"
Second Read step: "Turnaround"
Read wells
Full plate
Wavelengths
2 (450 nm, 630 nm)
Read Speed
Normal
Delay after plate movement
100 msec
Data Reduction
Define two Delta OD transformations (450-630
nm), one per Read data set
Synergy H1 Abs Test 3.prt
Parameter
Setting
Plate Type
96 WELL PLATE
Shake Step
Linear, 30 seconds, default frequency
Kinetic loop
Set a Run Time/Interval combination to read the
plate five times with minimal delay
Detection Method
Absorbance
Read Type
Endpoint
Optics Type
Monochromators
Read wells
A1..H6
Wavelength
340 nm
Read Speed
Normal
Delay after plate movement
100 msec
Synergy H1 Operator's Manual
Absorbance Testing | 79