Test Procedure
1. Prepare the
75% Test Solution
by mixing 15 mL of the 100% Test Solution with 5
mL of the PBS Solution.
2. Prepare the
50% Test Solution
by mixing 10 mL of the 100% Test Solution with 10
mL of the PBS Solution.
3. Carefully pipette the three solutions into a new 96-well microplate:
l
150 μL of the 100% Test Solution into all wells of columns 1 and 2
l
150 μL of the 75% Test Solution into all wells of columns 3 and 4
l
150 μL of the 50% Test Solution into all wells of column 5 and 6
4. Create a Gen5 experiment based on the
Synergy H1 Abs Test 3
protocol and read
the plate.
l
Save the experiment. Refer to the instructions on page
to perform calculations
and determine pass/fail.
l
Troubleshooting tips are provided on page
.
l
Test descriptions are provided on page
.
118 | Chapter 8: Instrument Qualification Procedures