Filter-based system, top optics:
The Detection Limit must be <=0.160 ng/mL to
pass.
o
Monochromator-based system, top optics:
The Detection Limit must be
<=0.160 ng/mL to pass.
Linearity Test
1. Calculate the Mean of the four wells for each concentration in columns 1-5.
2. Perform linear regression using these values as inputs:
Using Sodium Fluorescein
x
y
1000
Mean of the 1000 pM wells
500
Mean of the 500 pM wells
250
Mean of the 250 pM wells
125
Mean of the 125 pM wells
62.5
Mean of the 62.5 pM wells
Using Methylumbelliferone
x
y
100
Mean of the 100 nM wells
50
Mean of the 50 nM wells
25
Mean of the 25 nM wells
12.5
Mean of the 12.5 nM wells
6.25
Mean of the 6.25 nM wells
3. Calculate the R
2
value; it must be
>=0.9500
to pass.
Fluorescence Polarization (FP) Test
1. Using the raw data from the Parallel read:
o
Calculate the Mean Blank (wells A6–H6).
o
Calculate the Signal for each HPR well: Subtract the Mean Blank from its
measurement value.
o
Calculate the Signal for each LPR well: Subtract the Mean Blank from its
measurement value.
2. Using the raw data from the Perpendicular read:
o
Calculate the Mean Blank (wells A6–H6).
o
Calculate the Signal for each HPR well: Subtract the Mean Blank from its
measurement value.
Synergy H1 Operator's Manual
Fluorescence Testing | 99
Summary of Contents for Synergy H1
Page 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Page 2: ......
Page 26: ...BioTek Instruments Inc xxiv Preface ...
Page 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Page 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Page 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...