Figure 4-6
Connecting SHORT (0
W
) to the 42942A
Step 9.
Press the
SHORT [-]
key to start the short data measurement.
NOTE
When the short data measurement is completed, the softkey label changes to
SHORT [DONE]
.
Step 10.
Remove the SHORT (0
W
) standard from the 7-mm port of the 42942A. Then connect the
LOAD (50
W
) standard to the 7-mm port (the LOAD standard is furnished with the
42942A Terminal Adapter).