112
Chapter 5
Setting Measurement Conditions
Selecting Sweep Trigger/Measurement Point Trigger
Selecting Sweep Trigger/Measurement Point Trigger
You can select whether the event to be triggered (the trigger event) is a single sweep or a
measurement at a single point.
Step 1.
Press the
[Trigger]
key to display the trigger menu.
Step 2.
Press the
SOURCE [ ]
key to display the trigger source menu.
Step 3.
Press the
EVENT [ ]
key to select the event to be triggered (pressing the key toggles the
selection).
NOTE
If the internal trigger (free-run) is selected in “Selecting a Method to Start Measurement
(Trigger Source)” on page 111, the sweep specified in “Specifying Sweep Times and
Stopping Sweep” on page 114 is performed regardless of which trigger event is selected.
Note that a sweep takes a longer time when measurement point is selected for the trigger
event than when sweep is selected. This is because the instrument requires an additional
process to trigger measurement at each point.
Softkey label
Selected trigger event
EVENT [ON SWEEP]
Sweep (a single sweep for one time trigger)
EVENT [ON POINT]
Measurement point (a measurement at a single
point for one time trigger)
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......