388
Appendix B
Key Definitions
Softkeys displayed by pressing the [Source] key
Softkeys displayed by pressing the
[Source]
key
Table B-53
Source oscillator (Source Osc) menu
Key operation:
[Source]
Softkey
Function
GPIB command
FREQUENCY
Sets the fixed oscillator frequency (CW frequency) when performing sweeps
with a parameter other than frequency.
CWFREQ <value>
LEVEL
Sets the fixed oscillator level when performing sweeps with a parameter other
than oscillator level.
POWE <value>
OSC UNIT [ ]
Selects voltage [VOLT] or current [CURR] as the unit of oscillator level
configuration. The unit set here is applied to both fixed oscillator level and
oscillator level sweeps.
POWMOD
{VOLT|CURR}
BIAS MENU
ä
Displays Table B-54, “DC bias (Source Bias) menu,” on page 388.
(N/A)
Table B-54
DC bias (Source Bias) menu
Key operation:
[Source] - BIAS MENU
Softkey
Function
GPIB command
BIAS on OFF
Toggles on and off the dc bias output.
DCO {ON|OFF}
MEAS RANGE [ ]
ä
Displays Table B-55, “DC bias range (DC Bias Range) menu,” on page 388.
The currently selected dc bias range is displayed ([
<
1mA], [
<
10mA], or
[
<
100mA]) in [ ] of the softkey label.
(N/A)
MODE [ ]
ä
Displays Table B-51, “DC bias mode (DC Bias Mode) menu,” on page 387. The
currently selected dc bias mode is displayed (voltage source mode: [VOLT],
current source mode: [CURR], constant voltage source mode: [C-VOLT], or
constant current source mode: [C-CURR]) in [ ] of the softkey label.
(N/A)
VOLTAGE LEVEL
Sets the dc bias voltage level.
DCV <value>
CURRENT LEVEL
Sets the dc bias current level.
DCI <value>
MAX LIMIT VOLTAGE
Sets the voltage limit value (upper limit) when using the dc bias source in the
constant current mode. Within the range that does not exceed this upper limit
voltage, the constant current output of the dc bias operates.
MAXDCV <value>
MIN LIMIT VOLTAGE
Sets the current limit value (upper limit) when using the dc bias source in the
constant voltage mode. Within the range that does not exceed this upper limit
current, the constant voltage output of the dc bias operates.
MINDCV <value>
OSC MENU
ä
Displays Table B-53, “Source oscillator (Source Osc) menu,” on page 388.
(N/A)
Table B-55
DC bias range (DC Bias Range) menu
Key operation:
[Source] - BIAS MENU - MEAS RANGE [ ]
Softkey
Function
GPIB command
1mA
|
Selects < 1 mA (less than 1 mA) as the dc bias range
1
.
DCRNG M1
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......