Chapter 7
Setting Up the Display of Measurement Results
Displaying Phase Values without Wrapping at ±180
°
Displaying Phase Values without Wrapping at ±180
°
The Agilent 4294A provides a feature that allows you to display phase values without
wrapping them at +180
°
or -180
°
. This feature, called “expanded phase,” is useful when
measured phase values are in a continuous increase or decrease and some of them exceed
+180
°
or -180
°
. To set up the expanded phase feature, follow these steps:
Step 1.
Activate the trace for which you want to use the expanded phase feature.
NOTE
To expand the phase, your parameter must be either |Z|-
q
/ |Y|-
q
or COMPLEX Z-Y. You
should activate Trace B in the former case, and the trace whose axis format is polar in the
latter case.
Step 2.
Press the
[Format]
key to display the Format menu.
Step 3.
Press the
EXP PHASE on OFF
(or
EXP PHASE ON off
) key to turn on or off the expanded
phase feature (pressing the key toggles your selection).
To activate:
Press:
Trace A
[A]
Trace B
[B]
Softkey Labels
Description
EXP PHASE ON off
Indicates that the expanded phase feature is
enabled.
EXP PHASE on OFF
Indicates that the expanded phase feature is
disabled.
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......