![Agilent Technologies 4294A Operating Manual Download Page 419](http://html.mh-extra.com/html/agilent-technologies/4294a/4294a_operating-manual_2867884419.webp)
Appendix B
411
Key Definitions
Softkeys displayed by pressing the [System] key
B.
Ke
y De
fin
iti
on
s
DELTA LIMIT
Sets the limit center value of the measured value at the sweep stop point of the
segment
1
. When setting the center value, also set the difference value (difference
between the upper and lower limits) with the DELTA LIMIT key. This kind of
limit configuration set by the center value and the difference value can be used as
an alternative to the limit configuration set by the upper and lower limits.
LIMDSTOP <value>
more 2/3
Displays Table B-86, “Segment edit (Edit Limit) menu (3/3),” on page 411.
(N/A)
done
Has the same function as the done key in Table B-84, “Segment edit (Edit Limit)
menu (1/3),” on page 409.
LIMSDON
cancel
Has the same function as the cancel key in Table B-84, “Segment edit (Edit
Limit) menu (1/3),” on page 409.
LIMSQUI
1. Even if you have set the limits with the MIDDLE VALUE (center value) key and the DELTA LIMIT (difference value) key, you
can convert the table into the upper and lower limit format by pressing the UPPER LIMIT key or the LOWER LIMIT key. On
the other hand, even if you have set the limits with the UPPER LIMIT (upper limit) key and the LOWER LIMIT (lower limit)
key, you can convert the table into the center and difference value format by pressing the MIDDLE VALUE key or the DELTA
LIMIT key.
Table B-86
Segment edit (Edit Limit) menu (3/3)
Key operation:
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - more 1/3 - more 2/3
[System] - LIMIT TEST - EDIT LIMIT LINE - ADD - more 1/3 - more 2/3
Softkey
Function
GPIB command
TEST ON off
Toggles on and off the limit line test for the segment. When the limit line test is
off, * is displayed to the right of the segment number in the completed limit line
table. Note that * is not displayed when the pointer (>) is displayed to the right of
the segment number.
LIMSTEST {ON|OFF}
MKR
®
START
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the sweep start value of the segment.
MKRLIMSTAR
MKR
®
STOOP
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the sweep stop value of the segment.
MKRLIMSTOP
MKR
®
VALUE STAR
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the limit center point at the sweep start point of the segment.
MKRLIMVSTAR
MKR
®
VALUE STOP
Substitutes the sweep parameter value at the main marker on the current active
trace (A or B) into the limit center point at the sweep stop point of the segment.
MKRLIMVSTOP
more 3/3
Displays Table B-84, “Segment edit (Edit Limit) menu (1/3),” on page 409.
(N/A)
done
Has the same function as the done key in Table B-84, “Segment edit (Edit Limit)
menu (1/3),” on page 409.
LIMSDON
cancel
Has the same function as the cancel key in Table B-84, “Segment edit (Edit
Limit) menu (1/3),” on page 409.
LIMSQUI
Table B-85
Segment edit (Edit Limit) menu (2/3)
Key operation:
[System] - LIMIT TEST - EDIT LIMIT LINE - EDIT - more 1/3
[System] - LIMIT TEST - EDIT LIMIT LINE - ADD - more 1/3
Softkey
Function
GPIB command
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......