Chapter 7
169
Setting Up the Display of Measurement Results
Manual Scale Setting (for measurements other than COMPLEX Z-Y)
7. S
ett
ing
Up t
he Di
sp
la
y o
f
M
eas
ur
em
ent Res
ult
s
Figure 7-9
Setting the scale per division
Scaling the Trace Based on the Top and Bottom Values
Whether the vertical axis is linear or logarithmic, you can scale the trace by specifying the
highest (top) and lowest (bottom) values displayed within the graph.
Step 1.
Activate the trace you want to manually scale.
Step 2.
Press the
[Scale Ref]
key to display the Scale Reference menu.
Step 3.
Press the
TOP VALUE
key. The Parameter Setting field in the upper left area of the screen
indicates the top value currently in effect, that is, the value that corresponds to the line
displayed at the top of the graph.
Step 4.
Use the keys or rotary knob of the ENTRY block in one of the following ways to specify
the top value.
•
Enter the desired value with the numeric keys (
[0]
to
[9]
and
[.]
) and then press one of
the unit keys (
[G/n]
,
[M/
m
]
,
[k/m]
, or
[
´
1]
).
•
Turn the rotary knob (
m
m
m
m
) until the desired value is set.
•
Press the step keys ([
][
¯
]) to set the desired value.
To activate:
Press:
Trace A
[A]
Trace B
[B]
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......