126
Chapter 5
Setting Measurement Conditions
Setting Measurement Accuracy, Stability, and Time
Setting Measurement Accuracy, Stability, and Time
Setting measurement bandwidth
Step 1.
Press the
[Bw/Avg]
key to display the measurement bandwidth/averaging menu.
Step 2.
Press the
BANDWIDTH [ ]
key to display the measurement bandwidth setting menu.
Step 3.
Select the measurement bandwidth.
Step 4.
Press the
return
key to return to the measurement bandwidth/averaging menu. Note that
the measurement bandwidth currently selected is displayed in the key label
[ ]
of the
BANDWIDTH [ ]
key.
NOTE
Setting a greater measurement bandwidth will improve measurement accuracy, but keep in
mind that this requires a longer measurement time.
Averaging between sweeps (sweep-to-sweep averaging)
Sweep-to-sweep averaging computes each data point (vector value) based on an
exponential average of consecutive sweeps weighted by a user-specified averaging factor.
The sweep-to-sweep uses the algorithm shown in Equation 5-1.
Equation 5-1
Sweep-To-Sweep Averaging Algorithm
where:
Measurement bandwidth
Key stroke
1 (measurement speed maximized)
1 FAST
2
2
3
3
4
4
5 (measurement accuracy maximized)
5 PRECISE
Softkey label
Selected measurement bandwidth
BANDWIDTH [1]
1 (measurement speed maximized)
BANDWIDTH [2]
2
BANDWIDTH [3]
3
BANDWIDTH [4]
4
BANDWIDTH [5]
5 (measurement accuracy maximized)
A
n
S
n
F
-----
1
1
F
---
–
è
ø
æ
ö
A
n
1
–
´
+
=
Summary of Contents for 4294A
Page 1: ......
Page 2: ......
Page 4: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 16: ...8 ...
Page 30: ...22 Chapter1 Installation Power Cable Figure 1 2 Alternative Power Cable Options ...
Page 70: ...62 Chapter3 Front Rear Panel and LCD Display Items Displayed on the LCD ...
Page 436: ...428 AppendixB Key Definitions Softkeys displayed by pressing the Recall key ...
Page 454: ...446 AppendixC Error messages WRONG I O PORT DIRECTION ...
Page 468: ......